Document Type

Article

Publication Title

Journal of applied clinical medical physics [electronic resource] / American College of Medical Physics

Abstract

PURPOSE: This study aims to validate HDR brachytherapy dosimetry for cervical cancer patients utilizing the egs_brachy Monte Carlo (MC) simulation.

METHODS: Three cervical cancer patients treated with

RESULTS: TPS-calculated doses were greater than those obtained from MC simulations. For the CTV, the median percentage differences were 7.9% (Q1: 6.4%, Q3: 9.8%; range: 0.4%-10.4%) for D

CONCLUSION: CTV and critical organ doses calculated by the TPS were consistently greater than those obtained from MC simulations. This suggests that the TPS may overestimate dose distributions, especially in heterogeneous regions like the pelvis. These results emphasize the need for continued validation of TPS algorithms in HDR brachytherapy for cervical cancer.

First Page

70144

Last Page

70144

DOI

10.1002/acm2.70144

Publication Date

7-1-2025

Included in

Oncology Commons

Share

COinS