Discordance of insulin-like growth factor-1 results and interpretation on four different platforms.

Document Type

Article

Publication Title

Clinica chimica acta; international journal of clinical chemistry

Abstract

OBJECTIVE: Inconsistent Insulin-like Growth Factor 1 (IGF-1) measurements among different platforms have been observed. In this study, we compared the IGF-1 assay on four different platforms.

METHODS: A total of 110 serum specimens were analyzed in this comparison study. IGF-1 was measured on the three different chemiluminescent automated immunoassay of Siemens Immulite 2000 XPi, DiaSorin Liaison XL, IDS iSYS and LC-MS/MS method. Results were compared with Weighted Deming regression. Bias was evaluated using the Bland-Altman method.

RESULTS: Weighted Deming regression analysis showed approximately 36 % negative variation on Immulite, compared to Liaison (Immulite = 0.64 * DiaSorin + 2.95, r

CONCLUSIONS: There are wide variations between different platforms for IGF-1 measurement. The lack of standardization in IGF-1 measurement creates a challenge for clinicians to monitor IGF-1 and treat patients with pituitary disorders, when switching from one platform to another. The potential impact of the variations in IGF-1 measurement between different platforms should be taken into consideration when managing patients.