Early Safety Profile Using Pulsed Field Ablation: Prospective Multicenter DISRUPT-AF Study.
Document Type
Article
Publication Title
JACC. Clinical electrophysiology
Abstract
BACKGROUND: Pentaspline pulsed field ablation (PFA) is an emerging technology delivering ultra-short bursts of ablative high electrical field energy. There are limited studies on atrial fibrillation (AF) ablation safety using PFA in real-world experience in the United States.
OBJECTIVES: The study sought to evaluate safety in patients who underwent AF ablation with a pentaspline PFA catheter in a multicenter registry.
METHODS: Patients from the prospective, multicenter, nonrandomized, rigorously monitored, real-world DISRUPT-AF (A registry Based Collaborative To Measure Efficiency, Effectiveness, and Safety of Farapulse PFA Technology for AF) registry (NCT06335082), who underwent first-time AF ablation via PFA using the pentaspline catheter, from April 2024 to May 2025, across 20 centers in the United States, encompassing 77 operators, were included. Baseline characteristics and acute procedural safety outcomes were collected at 1- and 3-month follow-up based on the investigator's arm: acute, symptomatic monitoring only, or full monitoring. The primary safety outcome comprised procedure/device-related adverse events (AEs) after 1 and 3 months. Risk-adjusted analyses were conducted to assess AE predictors.
RESULTS: A total of 1,576 patients were included: the mean age was 68.0 ± 11.4 years, 38.6% were female, and mean body mass index was 31.5 ± 7.0 kg/m
CONCLUSIONS: In this multicenter, prospective U.S. registry, PFA demonstrated a low rate of procedure/device-related AEs. The absence of injury to the esophagus, phrenic nerve, and pulmonary veins highlights a tissue-selective profile that may represent a meaningful safety advantage over traditional thermal ablation, pending further direct comparisons.
DOI
10.1016/j.jacep.2026.04.032
Publication Date
6-23-2026
Recommended Citation
Wong EA, Zei PC, Hincapie D, Dai C, Koplan BA, Ruble S, Bezenek SM, Justice L, Natale A, Costello J, Dominic P, Campbell DN, Kessler D, Bode WD, Coffeen PR, Manogue M, Donnelly JA, Metzl MD, Wasserlauf J, Vivas Y, Eckart R, Thosani AJ, Cuoco F, Sidney DS, Mohanty GP, Belden WA, Friehling MS, Amin AK, Bailin S, Silverstein J, Cherian TS, Rajendra A, Meyers JD, Pelchovitz D, Day JD, Osorio J, Al-Ahmad A; DISRUPT AF Investigators. Early Safety Profile Using Pulsed Field Ablation: Prospective Multicenter DISRUPT-AF Study. JACC Clin Electrophysiol. 2026 Jun 23:S2405-500X(26)00454-8. doi: 10.1016/j.jacep.2026.04.032. Epub ahead of print. PMID: 42360264.